Explosion-proof slot resistance thermometer System V...f

Explosion-proof slot resistance thermometers are used to measure the winding temperature in the stator of generators or electric motors for the approved “intrinsic safety” and “increased safety” types of protection.

ExNWT’s can be equipped with platinum thin-film measuring resistors (chip) or also be manufactured as a wound version. The chip variant records the temperature at specific points in a very small area, since the temperature-sensitive length of the thin-film resistor is just a few millimeters. The coiled variant records the average temperature over the entire active length of the sensor.

Dimensions depend on the stator and can be adapted to the different slot sizes.

For inquiries, use our configurator or contact us directly.

Type ExNWT-f

Type ExNWT-s

Configure & Requests

Accuracy class according to IEC 60751

°C
DIN B
DIN A
(1/2 DIN B)
DIN AA
(1/3 DIN B)
0
± 0,3K
± 0,15K
± 0,1K
100
± 0,8K
± 0,35K
± 0,27K
200
± 1,3K
± 0,55K
± 0,44K
Scroll to Top